ShenzhenYijiajie Electronic Co., Ltd.

The company advocates the corporate culture of "integrity and dedication, pragmatic innovation, unity and cooperation, diligence and progress", pays attention to standardized management, advocates

Manufacturer from China
Verified Supplier
4 Years
Home / Products / Oxygen Gas Sensor /

CNLH Oxygen (O2) Gas Sensor Part Numbers: AA015-000 For Oxygen Analyzers 2-Electrode Electrochemical

Contact Now
ShenzhenYijiajie Electronic Co., Ltd.
Visit Website
City:shenzhen
Province/State:guangdong
Country/Region:china
Contact Person:MissXu
Contact Now

CNLH Oxygen (O2) Gas Sensor Part Numbers: AA015-000 For Oxygen Analyzers 2-Electrode Electrochemical

Ask Latest Price
Video Channel
Model Number :AA015-000 CNLH
MOQ :1
Payment Terms :L/C, D/A, D/P, T/T, Western Union, MoneyGram
Delivery Time :3-5work days
Measurement Range :0-2 ppm O2
Maximum Overload :Maximum Overload
Response Time (T95)* :< 20 seconds
Purge Time :< 6 hours
Recommended Load Resistor :10 Ω
Weight :Weight
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

CNLH Oxygen (O2) Gas Sensor Part Numbers: AA015-000 For Oxygen Analyzers 2-Electrode Electrochemical

Document Purpose:

The purpose of this document is to present the performance specification of the CNLH oxygen gas sensor.
This document should be used in conjunction with the Operating Principles (OP02) and the Product Safety Datasheet
(PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor
manufacture.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to
the Operating Principles (OP02).

* Specifications are valid at 20°C, 50% RH and 1013 mBar, using City Technology recommended
circuitry. Performance characteristics outline the performance of sensors supplied within the
first 3 months. Output signal can drift below the lower limit over time.

Operating Temperature Range -20°C to +50°C
Temperature Coefficient 0.2% signal/°C
Recommended Storage Temp 0°C to 20°C
Operating Pressure Range Atmospheric ± 10%
Operating Humidity Range 0 - 99% RH non-condensing
Long Term Output Drift <5% signal loss/year
Expected Operating Life See below
Storage Life 6 months in CTL container